| Output Voltage Ranges |
| Standard |
0-135 V, L-N
0-270 V, L-N |
| -HV Option |
0-156 V, L-N
0-312 V, L-N |
| -LV Option |
0-67.5 V, L-N
0-135 V, L-N |
| -EHV Option |
0-200 V, L-N
0-400 V, L-N |
| Output Frequency |
| See V-F derating for 17 Hz to 45
Hz performance |
| Models 751L, 1501L, 1503L,
2001L, 2750L-3, 4500L-3 and
6000L-3: |
17 Hz to 5 kHz |
Models 2750L-1, 4500L-1, 9000L
and 13500L: |
17 Hz to 2 kHz |
Models 6000L-1, 12000L and
18000L: |
17 Hz to 440 Hz |
| All models with -PT controller |
17 Hz to 550 Hz except 6000L,
12000L and 18000L |
| Total Harmonic Distortion |
| 45 Hz to 2 kHz |
1 percent |
| To 5 kHz |
2 percent |
| 50/60 Hz |
0.5 percent typical |
|
 |
|
 |
|
| Models 751L, 1501L, 1503L,
2001L |
115 V or 230 V single phase
47 Hz to 440 Hz |
| Models 2750L, 4500L, 9000L,
13500L: |
187 V - 252 V, L-L, 3 phase
342 V - 456 V, L-L, 3 phase with
-UP option |
| Models 6000L, 1200L, 18000L: |
208 V - 252 V, L-L, 3 phase.
2750L: 1 or 3 phase input 187 -
252 V |
|
 |
|
 |
|
| AC Noise Level |
160 mV rms typical |
| Connectors |
- Input provided on rear terminal
block
- Output provided on rear
terminal block (Remote sense
mating connectors are provided) |
| Protection |
- Overcurrent
- Overpower
- Short circuit
- Overtemperature
- Current limit trip standard with
programmable units. All units
have adjustable current limit |
|
 |
|
 |
|
 |
|
 |
|
| Programmability Choices |
| Specification |
Programmable -P |
Programmable -PT |
Manual -M |
| Controller Type |
Programmable Controller |
Fast Transient Controller |
Manual Control Oscillator |
|
| Voltage |
|
|
|
| Range |
0 - 135 V / 0 - 270 V L-N
Programmable range change Individual phase programming |
0 - 135 V / 0 - 270 V L-N
Programmable range change
Individual phase programming |
Variable pot control Switchable range change Option -RPV for 0-FS control
using 0 - 10 VDC input. |
| Accuracy |
± 0.135V from 5 V to 135 V
± 0.54 V from 135 V to 270 V
@ 25° C ± 1° C |
± 0.7% FS from 5 % to FS Constant line, load and temperature @ 25° C ± 1° C |
Variable control Analog meter readback |
| Load Regulation |
TRMS Sense: ± 0.05 % FS no load to full load |
- 0.5% FS from 45Hz to 100Hz
- 2% FS from 100Hz to 550Hz |
± 0.05% from 45Hz to 5kHz |
| Line Regulation |
± 0.02% FS for ± 10%
line change |
± 2% of full output for
a ± 10 % line change |
± 0.02 % of full output for
a ± 10 % line change |
| Stability |
± 0.015% FS over 24 hours at
constant line, load and temperature |
± 0.25% FS over 24 hours at
constant line, load and temperature |
± 0.015% FS over 24 hours at
constant line, load and temperature |
| Initial Value |
0.0 or 5.0 Vrms field selectable |
0 Vrms |
N/A |
| Setting Time |
16 msec, no-load from 5 V to within 2 % of final value; 16 msec, full load from
5 V to within 15 % of final value |
0.5 msec |
N/A |
| Programmable THD |
N/A |
0 - 20% THD clipped sine
1% resolution |
N/A |
| Amplitude Modulation |
N/A |
0 to 5V RMS generates 0 to 11% amplitude modulation of output voltage. 45 Hz to 5 kHz input |
0 to 5 V RMS generates 0 to 11 % amplitude modulation of output voltage. |
| Frequency |
|
|
|
| Range |
2750L-1P, 4500L-1P and all multibox systems:
45 Hz to 2 kHz
-3P and 751L - 2001L:
45 Hz to 5 kHz
6000L, 12000L and 18000L:
45 Hz to 440 Hz |
6000L, 12000L and 18000L:
45 to 440 Hz
All other models:
45 Hz to 550 Hz |
2750L-1M, 4500L-1M
and all multibox systems:
45 Hz to 2 kHz
-3M and 751L - 2001L:
45 Hz to 5 kHz
6000L, 12000L and 18000L:
45 Hz to 440 Hz |
| Resolution |
0.01 Hz; 45.00 Hz to 99.99 Hz
0.1 Hz; 100.0 Hz to 999.9 Hz
1 Hz; 1000 Hz to 5000 Hz |
0.01 Hz; 45.00 Hz to 99.99 Hz
0.1 Hz; 100.0 Hz to 550.0 Hz |
3 digits |
| Accuracy |
± 0.001 % of programmed value |
± 0.001 % of programmed value |
± 0.005 % of set value |
| Initial value |
Any within range |
Any within range |
Setting |
| External Sync Input |
TTL level |
TTL level |
N/A |
| Phase |
|
|
|
| Range |
Phase B and/or C relative to phase A: 0 to ± 360° in 0.1° increments |
Phase B and/or C relative to phase A: 0 to ± 360° in 0.1° increments |
N/A |
| Accuracy |
± 2 ° |
± 2 ° |
± 3 ° |
| Current |
|
|
|
| Programmable Limit |
Adjustable trip |
Adjustable trip |
Adjustable foldback with
recovery |
| Remote Inhibit |
Contact closure turns output off |
Contact closure trips unit off. Sets defaults. |
Contact closure turns output off |
| Measurements |
|
|
|
| Voltage |
resolution 0.1 Volt, accuracy ± 10 digits |
N/A |
| Current |
resolution 0.01 Amp or 0.1 Amp, accuracy ± 10 digits |
N/A |
| Power |
resolution 1 W or 0.01 kW, accuracy ± 10 digits |
N/A |
| Phase Angle |
resolution 0.1°, accuracy ± 2° to 2 kHz, ± 3° to 5 kHz |
N/A |
| Power Factor |
range 0.000 to 0.001 |
N/A |
| Frequency |
resolution four decades, accuracy ± 0,02 Hz to 99.99 Hz,
± 0.2 Hz to 500.0 Hz, ± 0.5 Hz to 999.9 Hz, ± 10 Hz to 5 kHz |
N/A |
| Apparent Power |
resolution 1 VA or 0.01 kVA, accuracy ± 10 digits |
N/A |
|
 |
|

|
|
 |
| Plug-in Controller Concept Provides Choice of Features |
|
| All L-Series models offer a choice of
three controllers, allowing you to specify
the levels of performance and control
best suited to your requirements. |
| Type -P |
the standard programmable
controller, uses True RMS sensing, providing
the most accurate output voltage
regulation. Output settling times for
the -P controller are longer than the
-PT controller due to the RMS sense
response time. With Type -P, transients
are programmable over time or cycles. |
| Type -PT |
uses a fast real-time servo,
instantly creating exact waveform definitions
ideal for applications such as
switching DC power supplies where
real-time feedback and fast output settling
times are critical. The -PT controller
is recommended for applications that require fast transients to be programmed
over time or cycles. |
| Type -M |
the manual controller, is
ideal for portable or benchtop applications
where local control is sufficient.
An optional remote programmable voltage
input can be added to allow amplitude
control using a DC input signal. |
| Single and Three Phase Versions |
All controllers are available in either
single or three phase versions. For special
applications, two phase or split
phase configurations can be ordered as
well. For three phase -P and -PT controllers,
a phase mode option can be
added which allows switching between
both single and three phase output
modes without the need to rewire the
output terminals. |
| Measurements |
Both -P and -PT controllers provide
a full range of output readback measurements
, either via front-panel display
or over the standard IEEE-488 bus.
Measurements provided are Volt RMS,
Current RMS, Power, Apparent Power,
Power Factor, Frequency and Phase. |
| Controller and Amplifier Options |
The L Series is highly configurable
using a wide array of options for both
the amplifier and the controller. This
makes the L Series one of the most
versatile AC power solutions on the
market. If your application requirements
can not be met using any of the options
listed here, contact the factory for
configuration assistance. |
|
 |
|
 |
|
 |
|
 |
|
 |
 |
|